LM OG-AM series
Oval gear meters with MID approved register and integrated
datalogger
LM OG-AM
LM OG-TAERM
LM OG-CNDAM
User Manual
LM_OG-AM_BA_02_1163 (May 2019)

Advertisement

Table of Contents
loading

Summary of Contents for Badger Meter Oval Gear LM OG-AM Series

  • Page 1 LM OG-AM series Oval gear meters with MID approved register and integrated datalogger LM OG-AM LM OG-TAERM LM OG-CNDAM User Manual LM_OG-AM_BA_02_1163 (May 2019)
  • Page 2: Table Of Contents

    LM OG-AM series CONTENT 1. Disclaimer ..............................................1 2. Basic safety recommendations ......................................1 3. Before putting into operation ......................................2 3.1 Recommended filter sizes ......................................3 4. Warranty ..............................................3 5. Details of unit operation LM OG-AM series ................................... 4 5.1 RESET button ..........................................
  • Page 3: Disclaimer

    However, no responsibility is assumed for inaccuracies, nor does Badger Meter assume any liability arising out of the application and use of the equipment described. Should the equipment be used in a manner not specified by Badger Meter, the protection provided by the equipment may be impaired and the warranty voided.
  • Page 4: Before Putting Into Operation

    Before putting into operation Repair of faults Disconnect all units from power supply and have it repaired by a qualified service person if any of the following occurs: WARNING Failure to adhere to these safety instructions may result in damage to the product or serious bodily injury.
  • Page 5: Recommended Filter Sizes

    WHATSOEVER INCLUDING BUT NOT LIMITED TO IMPLIED WARRANTIES (EXCEPT OF TITLE) OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Badger Meter shall not be liable for any defects attributable to acts or omissions of others after shipment, nor any consequential, incidental or contingent damage whatsoever.
  • Page 6: Details Of Unit Operation Lm Og-Am Series

    Details of unit operation LM OG-AM series 5. DETAILS OF UNIT OPERATION LM OG-AM SERIES 5.1 RESET button The display of the resettable totalizer (batch register) can be put to zero by pushing the RESET button. A RESET is not possible during a batch process.
  • Page 7: Totalizer

    Details of unit operation LM OG-AM series 5.6 Totalizer The total dispense quantity is shown in the bottom line. The factory programmed correction factor results from accuracy testing of bare meter with the mounted electronic register: Pressing the TOTAL and the RESET button simultaneously for 3 seconds will show the programmed correction factor.
  • Page 8: Programming Of Lm Og-Am And Lm Og-Taerm Series

    Programming of LM OG-AM and LM OG-TAERM series 6. PROGRAMMING OF LM OG-AM AND LM OG-TAERM SERIES The units of measure and the correction factor can be configured in the programming mode. A program button on the backside of the register will enable the mode, only accessible when the register has been removed from the meter. Removing the register from the meter will break the sealing.
  • Page 9: Calculating The Correction Factor

    Calculating the correction factor 7. CALCULATING THE CORRECTION FACTOR Two different ways are described below to eliminate the existing deviation of a meter. The reference quantity (“quantity dispensed”) should always be measured with an appropriate measuring device. • Approved piston or bell prover •...
  • Page 10: Browse And Program Data Logger

    Browse and program data logger 9. BROWSE AND PROGRAM DATA LOGGER Pushing the RESET will automatically save the current dispense in the datalogger, if the quantity is bigger than 0,49 liters. The datalogger is capable to store 500 dispenses, any additional transaction will overwrite the oldest one without notice. 9.1 Search for a dispense The sample below shows how to search for a dispense on a certain day.
  • Page 11: Programming The Datalogger

    Display correction factor 9.2 Programming the datalogger Below you will find the procedure to change time and date. The date can only be changed with the program button on the backside, which is only accessible when the register has been removed from the meter. Removing the register from the meter will break the sealing.
  • Page 12: Checksum & Serial Number

    Checksum & serial number 11. CHECKSUM & SERIAL NUMBER Pressing RESET for 3 seconds will display the checksum and serial number in the bottom line. 3A33.07 3A33 = Checksum over the MID relevant source code Checksum = Sequential version number, not MID relevant AC001 TOTAL = The year of production...
  • Page 13: Mid Approval

    MID approval 14. MID APPROVAL You will find all approvals applying to this meter series under www.badgermeter.de/en/downloads/approvals/ 15. MANUFACTURER DECLARATION You will find all CE declarations applying to this meter series under www.badgermeter.de/en/downloads/ce-declarations/ 16. DIN ISO AND EMC CERTIFICATE Please refer to our valid DIN ISO and EMC certificate under www.badgermeter.de/en/downloads/certificates/ 17.
  • Page 14 Return of goods for repair / Harmlessness declaration Page 12 LM_OG-AM_BA_02_1163 May 2019...
  • Page 15 Return of goods for repair / Harmlessness declaration May 2019 LM_OG-AM_BA_02_1163 Page 13...