Invacare® Perfecto₂™ Oxygen Concentrator

5.7 Preventive Maintenance Checklist

Model No:
ON EACH INSPECTION
Record Date of Service
Record Elapsed Hours on
Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min.
Flow Rate
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
UNITS WITH SensO
— EVERY 26,280 HOURS
2
Check Oxygen
Concentration*
Clean/Replace Cabinet Filter
Check/Replace Outlet HEPA
Filter*
Check/Replace Compressor
Inlet Filter*
Check Power Loss Alarm
26,820 hours are equivalent to usage 24 hours per day, 7 days per week for 3 years.
*To be conducted by provider or qualified service technician. Refer to 8.6 Oxygen Sensor, page 44.
16
Serial No:
1154245-F-02