Diameter of
Distance
Target Reflector
A(dB)
(mm)
(d)(mm)
50
44.5
75
44.5
1.58
A=44.5dB
@3MHz
100
44.5
200
44.5
50
42.0
75
42.0
2.38
A=42.0dB
@3MHz
100
42.0
200
42.0
Doppler Frequency (Hz)
SONOTRAX(2009) Series Ultrasonic Pocket Doppler User Manual
Two-way Attenuation
B=∑B
+B
a
w
∑B (T: ultrasonic attenuation
phantom No.
B
:dB)
a
T
6#
3#
-
-
B
43.6
13.5
-
-
a
T
6#
3#
-
-
B
43.6
13.5
-
-
a
T
6#
3#
-
-
B
43.6
13.5
-
-
a
T
6#
3#
-
-
B
43.6
13.5
-
-
a
T
6#
3#
-
-
B
43.6
13.5
-
-
a
T
6#
3#
-
-
B
43.6
13.5
-
-
a
T
6#
3#
-
-
B
43.6
13.5
-
-
a
T
6#
3#
-
-
B
43.6
13.5
-
-
a
500
V
(r.m.s)
V
S
n
B
B
mV
w
(dB)
(dB)
0
57.1
112.3
52.44
0
57.1
108.4
52.28
0
57.1
113.8
54.56
0
57.1
112.2
54.82
0
57.1
109.0
53.46
0
57.1
113.8
52.43
0
57.1
110.4
54.35
0
57.1
112.7
54.46
Velocity of Target
(cm/s)
107
V
r
.
m
. .
s
 
 
s
C
20
log
(r.m.s)
10
V
r
.
m
. .
s
mV
n
dB
6.61
6.34
6.39
6.22
6.18
6.73
6.16
6.32
12.5
Overall
Sensitivity
(S=A(d)+B+C)
dB
108.2
107.9
107.9
103.3
105.2
105.8
105.2
105.4