SONOTRAX(2009) Series Ultrasonic Pocket Doppler User Manual
PD
(µ sec)
PRF
(Hz)
EBD
Operating
Control
Fixed
Conditions
System: SONOTRAX
Transducer: CD3.0
Acoustic Output
Global Maximum Value
Az.
(cm)
Ele.
(cm)
Acoustic Output Reporting Table for Track 1
Non-scanning Mode
97
CW
N/A
2.50
1.25
Operating Mode: CW mode
Working Frequency:
I
SPTA.3
MI
(mW/cm 2 )
0.0041
CW
N/A
3.0 MHz
I
SPPA.3
(W/cm 2 )
1.63