Ecg Cannot Be Previewed With Seer Light Connect Device - GE SEER Light Service Manual

Ambulatory recorder/controller
Table of Contents

Advertisement

ECG Cannot Be Previewed with SEER Light Connect Device

2-10
SEER Light/SEER Light Extend Recorder: Troubleshooting
ECG cannot be previewed
with the SEER Light
Connect Device.
Is the USB cable
from the Connect
plugged in?
Yes
Is the orange LED
on the Connect on?
Yes
Is the distance from
the controller within
4.9 ft. (1.5 m)?
Yes
Is the I.R.
window facing the
controller properly?
Yes
Were the batteries
installed within one
hour?
Yes
Has recording
already started?
No
Call GE service.
SEER Light Ambulatory Recorder/Controller
2019818-008
No
Plug it in.
No
Call GE Service.
No
Position the controller
within 4.9 ft (1.5 m).
No
Reposition the I.R. window
properly, or remove any
obstacle between the
recorder and the controller.
No
Reinstall the batteries.
Yes
Cannot preview ECG by I.R.
operation if more than 3
minutes has elapsed since
recording started.