Cannot Detect Pacemaker Pulse
2-20
SEER Light/SEER Light Extend Recorder: Troubleshooting
Cannot detect
pacemaker pulse.
Insert the batteries while
pressing the STOP button.
Disconnect the patient cable
and touch the electrode
connector with a finger.
Does audible beep
sound?
No
Call GE Medical Systems
Information Technologies
Service.
SEER Light Ambulatory Recorder/Controller
2019818-008
Yes
Normal pacemaker detection circuit.
Confirm pacemaker pulse detection
by the Preview function of the
controller. Reposition the CH1 if the
pulse signals are too small.
Revision B