Recording Cannot Start with the Controller (I.R.)
Revision B
SEER Light/SEER Light Extend Recorder: Troubleshooting
Recording cannot start
with the controller (I.R.).
Does the controller
display "untransmitted
data?"
No
Does
the controller display
"Already started?"
No
Is the distance
from the controller
within 4.9 ft
(1.5 m?)
Yes
Does the I.R. window
face the controller
properly?
Yes
Call GE Service.
SEER Light Ambulatory Recorder/Controller
2019818-008
Yes
Transfer previous data or
press the START button
when the controller displays
"Data still remains."
Yes
Reset the batteries over 30
seconds after removing them.
No
Adjust the position
within 4.9 ft. (1.5 m)
from the controller.
No
Reposition the I.R. windows
to face each other properly.
Remove any obstacles
between the recorder and
the controller.
2-11