Operating Instructions
2.
Calibration
(depending on the selection in system 7.2 Sensors)
2.1
Calibration factor
2.2
Calibration filter
2.3
Calibration points
(support points)
2.4
Calibration
2.4.1
P1 value
(2.4.2) P1 calibration
. . . (depending on the number of support points)
2.4.n
Pn value
(2.4.n) Pn calibration
input: 0.01 ... 9.99
Input: n = 1 ... 9999
Input: 2 ... 5
Calibration sub-menu
Input: Measurement
Transfer: Raw value
Input: Measurement to be displayed
Transfer: Current raw value
Factor for the subsequent adjustment of the
actual measurement. All measurements
are measured with this factor.
Average number n for the raw value when
performing a calibration. n values are
combined into an average value with PT1
character, in order to obtain a more stable
representation.
This filter also affects the representation of
the calculated raw value in the online
window.
Number of support points for a
linearisation above the operating range.
Output measurement in the selected
mass/time unit.
Transfer of the current raw value (filtered)
from the mass flow with the key [ ← ].
The value can also be entered directly.
For additional support points (depending on
[ 2.3 ]), additional value pairs can be set.