D. Check limit switches for functionality by using a finger and voltmeter.
- The micro switch up is normally open (no signal) and when activated,
by hook or tilted band, it should give a signal (closing) and stop winding movement.
o Bad switch should be changed.
- The micro switches down is normally closed (signal) and when closed, by end of
strap (no winding left on drum), it should give no signal and stop winding movement.
o Bad switch should be changed.
- The micro switch pre tension is normally open (no signal) and when activated,
by tension on the lift strap, it should give a signal and allow winding.
o Bad switch should be replaced.
When checking switches there might also be need for adjustment, especially for the micro switch down. Ensure that
there are at least 2 laps, overlapping, of strap on the drum before micro switch down breaks the winding movement.
More laps results in decreased lifting height of RiseAtlas.
Above actions and measurements covers most of any potential electronic issues.
Also check that emergency switch moves correctly.
Mechanical movements are checked by running the motor and check:
1. Motor
2. Transmission gears.
3. Single - Fault gear.
All screws and bolts should be tightened. If there is no positive result of the advanced trouble shooting, please contact
service support.

6. Resetting the service indicator

Only authorized technicians shall perform this operation on RiseAtlas and thorough understanding of the product is a
must. Access inside RiseAtlas is described in chapter 3 and necessary tools are described in chapter 7.
The service indicator can be reset via hand control or via PCB-card.
Resetting the service indicator via hand control (need for 6-button hand control),
Ensure that power of the RiseAtlas is ON and then press simultaneously ON-button and both SIDE-buttons (all together
3 buttons) for 5 seconds. This will reset the service indicator.
Resetting the service indicator via PCB-card,
Start the operation by Access inside RiseAtlas, chapter 3.
Ensure that power of the RiseAtlas is ON and then press the red button
for 11 seconds. This will reset the service indicator.