SECTION 4 – TECHNICAL DATA
Table A1 - Ultra-Low Variable Energy (150 VE) Responder AED Waveform
Patient's
Impedance
(Ohms)
25
50
75
100
125
150
175
Table A2 – Low Variable Energy (200 VE) Waveform Responder AED Waveform
Patient's
Impedance
(Ohms)
25
50
75
100
125
150
175
Page 32 of 35
Figure A1. STAR BIPHASIC WAVEFORM
Phase 1
Voltage*
Duration*
(Volts)
(ms)
1393
3.3
1420
4.5
1430
5.8
1434
7.0
1437
8.3
1439
9.5
1441
10.8
Phase 1
Voltage*
Duration*
(Volts)
(ms)
1609
3.3
1640
4.5
1651
5.8
1656
7.0
1660
8.3
1662
9.5
1663
10.8
© 2010 General Electric Company
Phase 2
Voltage*
Duration*
(Volts)
(ms)
743
3.2
909
3.2
973
3.2
1007
3.2
1027
3.2
1040
3.2
1049
3.2
Phase 2
Voltage*
Duration*
(Volts)
(ms)
858
3.2
1050
3.2
1124
3.2
1163
3.2
1186
3.2
1201
3.2
1212
3.2
Energy**
(Joules)
145-196
128-173
116-156
108-146
102-138
98-132
95-128
Energy**
(Joules)
193-260
170-230
155-209
144-194
136-184
131-176
126-170