SECTION 4 – TECHNICAL DATA
EMISSIONS
Field
E-M
Magnetic
IMMUNITY
Field
E-M
Magnetic
ESD
ENVIRONMENTAL CONDITIONS
Field
Free Fall Drop
Bump
Vibration
(Random)
Vibration (Sine)
Enclosure
Protection
SHIPPING AND TRANSPORT CONDITIONS
ISTA Procedure 2A
Page 30 of 35
Standard or Compliance
IEC 55011/CISPR 11, Group 1, Class B
ANSI/AAMI DF39, <0.5mT on surface, except for within
5cm of the lid magnet and the speaker
Standard or Compliance
IEC 61000-4-3, Level X, (20V/m)
IEC 60601-2-4, Section 36.202.3 (20-V/m)
AAMI DF39, Section 3.3.21.2.1
IEC 61000-4-8 (2001)
IEC 60601-2-4 (2002), Section 36.202.8
AAMI DF39, Section 3.3.21.2.3 80A/m, 47.5Hz – 1,320Hz
IEC 61000-4-2, Level 3
IEC 60601-2-4 (2002), Section 36.202.2
6KV contact discharge, 8KV air gap discharge
Standard or Compliance
IEC 60068-2-32 (1975) AM 2 (1990), 1 meter
IEC 60068-2-29 (1987), 40g and 6000 bumps
IEC 60068-2-64 (1993): 10Hz –2KHz, 0.005 – 0.0012
g2/Hz
IEC 60068-2-6 (1995): 10Hz – 60Hz, 0.15 mm and 60Hz
– 150 Hz, 2g
IEC 60529 (2001), IP24
© 2010 General Electric Company