SECTION 4 – TECHNICAL DATA
These protocols are selected by using our ServiceLink software program. The five biphasic energy protocols
available are as follows:
Energy Protocols
Factory Default
Protocol #2
Protocol #3
Protocol #4
Protocol #5
1
The Ultra-Low Energy (150 VE), Low Energy (200 VE) and High Energy (300 VE) shocks are variable energy. The
actual energy is determined by the patient's impedance.
2
Allowable energy range
Page 34 of 35
1
Shock Sequence
1.
2.
3.
1.
2.
3.
1.
2.
3.
1.
2.
3.
1.
2.
3.
© 2010 General Electric Company
Energy Level (VE)
200
300
300
200
200
300
150
200
200
150
150
200
200
200
200
2
Energy Range
(J)
126-260
170-351
170-351
126-260
126-260
170-351
95-196
126-260
126-260
95-196
95-196
126-260
126-260
126-260
126-260
2022100-201 Rev C