Aligning the Camera to the DRS Laser Sensor

To align the video camera on your Cobra system to the DRS laser sensor, follow
the steps below. You will need a 2 mm hex key (Allen) wrench.
1.
2.
On-line Help: See the topic, How to Focus the Sensor, for more information.
3.
4.
Focusing and Aligning the Camera
Power up your Cobra system and access Scan-X.
Focus the sensor over the surface of a part, and observe the laser spot on the
LCD display.
-
If the laser spot appears centered in the LCD display, stop here. You do
not need to perform the rest of this procedure.
-
If the laser spot does not appear centered in the LCD display, continue
with Step 3 of this procedure.
Use a 2 mm hex key (Allen) wrench to remove the four screws securing the
camera cover to the camera assembly (see Figure 2-4 on page 2-6).
Align the camera by moving it slightly, as described below, while observing
the position of the laser spot on the LCD screen. Continue adjusting the
horizontal and/or vertical adjustment screws until the laser spot appears
centered in the display. Use a 2 mm hex key (Allen) to adjust the screws.
-
To move the laser spot to the right, tighten the horizontal adjustment
screw (see Figure B-2 on the following page)
-
To move the laser spot to the left, loosen the horizontal adjustment
screw (see Figure B-2 on the following page)
-
To move the laser spot up, loosen the vertical adjustment screw (see
Figure B-2 on the following page)
-
To move the laser spot down, tighten the vertical adjustment screw (see
Figure B-2 on the following page)