Swapping Sensors on Systems Not Equipped
with the Video Camera Option
This subsection only applies to Cobra systems not equipped with the video
camera option, and describes how to remove and replace sensors.

Removing a Sensor

To remove the sensor currently installed on your Cobra system, follow the steps
below. You will need a 3 mm hex key (Allen) wrench.
1.
Never connect or disconnect the laser sensor cable from the sensor while power
is applied to Cobra. Doing so could seriously damage the sensor.
2.
Hold onto the sensor as you remove the three sensor mounting screws. If the
sensor is not supported, it can fall off the sensor mounting pad and be damaged.
3.
4-8
Power down your Cobra system.
Disconnect the laser sensor cable from the top of the sensor, see Figure 4-7.
Hold the sensor with one hand and remove the three M4 socket-head cap
mounting screws with your free hand. Use a 3 mm hex key (Allen) wrench.
Figure 4-7. Removing the Laser Sensor
Swapping Sensors