Section 5
This section describes how to adjust the rotational and vertical mounting
position of the sensor mounting pad (Z slide). It may be necessary to adjust the
rotational mounting position of the mounting pad in the event of a certification
failure and any time the vertical position of the mounting pad is altered. The
vertical position of the mounting pad may need to be altered to fit a specific
application in the event that the slide cannot be moved into the desired position
using the Focus knob.
Note: If the mounting position of the sensor mounting pad (Z slide) is altered in
any way from the factory installed position, your Cobra should be re-certified to
ensure system accuracy. See the topic, System Certification, in Section 2 for
more information. Consult your local authorized sales representative for more
information on field certification of your Cobra system.

Rotational Alignment

The rotational alignment procedure aligns the sensor mounting pad so it is
square to the working surface. Before performing this procedure, contact your
local authorized sales representative to determine whether or not alignment is
necessary. This procedure requires that Cobra is installed and level.
Note: The sensor mounting pad was pre-aligned and your system was certified
at the factory (see the topic, System Certification, in Section 2 for more
information). The following procedure will disrupt the alignment and void the
certification of your Cobra system.
To rotationally align the sensor mounting pad follow the steps below.
1.
2.
Adjusting the Sensor Mounting Pad
Remove the DRS laser sensor, as described in Section 4.
Loosen the Locking knob (see Figure 5-1).
Adjusting the
Sensor Mounting Pad
5-1