Programming the transmitter
6.6.3

Procedure - configuring temperature related settings

Complete the following steps to set the temperature units and to select automatic or
manual temperature correction.
1.
2.
3.
4.
5.
6.
42
measurement and calibration temperatures differ by no more than
about 2 °C. Manual temperature correction is useful if the sensor
temperature element has failed and a replacement sensor is not
available.
Press MENU.
The main Menu screen appears.
Move the cursor to Program and press ENTER.
Choose Temperature.
The screen summarizes the present sensor configuration. There are between two
and three items. Units and S1 Temp Comp always appear. If manual temperature
compensation was selected, the manual temperature value entered for the sensor
(S1 Manual) also appears.
To make a change, move the cursor to the desired line and press ENTER.
A screen appears in which the present setting can be edited.
To store a setting, press ENTER.
For an explanation of terms, see
To return to the main display, press MENU and then EXIT.
Section
6.6.2.