Connection Of Feed Tubes - Hach TitraLab User Manual

Total hardness content analyzer
Hide thumbs Also See for TitraLab:
Table of Contents

Advertisement

Installation
Figure 7
Position of protective stickers
3.4.3.2

Connection of feed tubes

Figure 8
Connection of feed tubes
18
The system is delivered with two feed tubes with bottle closures (DIN 45 thread). These
feed tubes are supplied pre-installed on the TitraLab and only need to be connected to the
relevant reagent bottles.
1. Connect the tube from the IN socket of the burette valve to the titrant bottle (EDTA
0.010 M TitraVer, P/N 205-53). Screw on the bottle closure to do this.
2. Connect the tube from the IN socket of the pump to the acid bottle (Hardness Buffer
reagent, TRIS 0.1 M). Screw on the bottle closure to do this.
3. Place a desiccant cartridge into the designated opening in each bottle closure.