Service, cleaning, and maintenance

The ICS Impulse System equipment does not require preventive maintenance. Observe the following recommended
guidelines regarding cleaning and maintenance.
Keep the instrument clean and as free of dust as possible. Remove dust using a soft cloth or brush.
If required, clean the goggle housing using a damp cloth moistened with a mild detergent and water solution. Do not
allow any moisture to get inside the goggles. Keep moisture away from the cooling fan.
Caution
Never spray or immerse the goggle components with the cleaning solutions. This could contaminate the
electronics and/or optics.
If required, clean the mirror and/or the camera lens using the supplied cleaning cloth. The presence of fingerprints on
the mirror surfaces could cause inaccurate pupil detection. If the camera or mirror is not clean, artifacts will be seen
during data collection.
Caution
Improper cleaning of either the mirror or the camera lens may scratch the surfaces.
Caution
A small cooling fan is inside the goggles housing. Keep the fan blades free from any obstruction.
Replacing the strap
1. Remove the face cushion.
2. Use a pen to push the plastic clip down and pull out the strap clip attached to the goggle.
Otometrics - ICS Impulse USB
Cable clip
Carrying case
Wall mount
User Guide
8-35-36900
8-35-36700
8-62-45600
53

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ICS Impulse USB and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF