Natus Medical Denmark ICS products are not designed to be used in conjunction with any devices not approved by Natus
Medical Denmark. Summation of combined unapproved parts could result in increased electrical leakage. All parts of the
ICS Impulse System are suitable for use within the patient environment.
When the ICS Impulse System is used in conjunction with a test device, make sure that all information, cautions, and warn-
ings in the manual for the test device are followed.
For safety specifics concerning test modules and test devices, see the specific manuals.
1. There are no user-serviceable parts inside the ICS Impulse System goggles. For the sake of safety and in order not to
void the warranty, service and repair of electro-medical equipment should be carried out only by the equipment man-
ufacturer or by service personnel at authorized workshops. In case of any defects, make a detailed description of the
defect(s) and contact your supplier. Do not use a defective device.
2. Keep the unit away from liquids. Do not allow moisture inside the unit.
3. Do not use the instrument in the presence of flammable anesthetics (gases).
4. A Class 2 Laser product is used for calibration. The laser beam projects from the front of the goggles onto a solid sur-
face.
Caution
other than those specified herein, may result in hazardous radiation exposure.
5. It is recommended to install the unit in an environment that minimizes the amount of static electricity. For example,
anti-static carpeting is recommended.
6. No parts may be eaten, burnt, or in any way used for purposes other than the applications defined in the Intended Use
section of this manual.
7. The device can be disposed of as normal electronic waste, according to local regulations.
8. For safety reasons, accessories connected to the equipment's outlet fittings must be identical to the type supplied
with the system.
9. The device is disconnected from the mains by removing the USB cable.
10. Do not touch non-medical parts, such as the laptop/computer or printer and the patient at the same time.
11. Exposure to electromagnetic fields can result in interference with the process of recording correct measurements. ICS
Impulse System cameras and gyroscopes are sensitive to electrical disturbances. Avoid static discharges and elec-
tromagnetic fields.
12. No defibrillators or high-frequency surgical equipment should be applied to the patient when connected to the ICS
Impulse System at any time.
13. Immediately discontinue use of the equipment if skin irritation or discomfort occurs.