Recommended separation distances between portable and mobile RF
communications equipment and the EBP-UA5 Upper-Arm Blood Pressure Monitor
The EBP-UA5 Upper-Arm Blood Pressure Monitor is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the EBP-UA5 Upper-Arm Blood Pressure Monitor
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the EBP-UA5 Upper-Arm Blood Pressure Monitor as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter W
0.01
0.1
1
10
100
26
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
N/A
N/A
N/A
N/A
N/A
80 MHz to 800 MHz
0.035
0.11
0.35
1.11
3.5
80 MHz to 2.7 GHz
0.07
0.22
0.7
2.21
7