Guidance and Manufacturer's Declaration – Electromagnetic Immunity –
for all EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
3 V rms
150 kHz to 80 MHz
Conducted RF
6 V in ISM bands
IEC 61000-4-6
between 0,15 MHz and
80 MHz
Radiated RF
10V/m 80 MHz to 2.7
IEC 61000-4-3
GHz
24
Compliance Level
Portable and mobile RF communications equipment should
be used no closer to any part of the EBP-UA5 Upper-
Arm Blood Pressure Monitor, including cables, than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
N/A
Recommended separation distance:
where P is the maximum output power rating of the
10V/m 80 MHz to 2.7
transmitter in watts (W) according to the transmitter
GHz
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey, should be less than the
compliance level in each frequency range.
Electromagnetic Environment - Guidance
80 MHz to 800 MHz
800 MHz to 2.7 GHz