ETEKCITY EBP-UA5 User Manual page 22

Upper-arm blood pressure monitor
Hide thumbs Also See for EBP-UA5:
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration – Electromagnetic Immunity –
The EBP-UA5 Upper-Arm Blood Pressure Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the EBP-UA5 Upper-Arm Blood Pressure Monitor should assure that it is used in such an
environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
22
for all EQUIPMENT and SYSTEMS
IEC 60601 Test Level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV differential mode
Compliance Level
Floors should be wood, concrete, or
± 8 kV contact
ceramic tile. If floors are covered with
± 15 kV air
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that
N/A
of a typical commercial or hospital
environment.
Mains power quality should be that
N/A
of a typical commercial or hospital
environment.
Electromagnetic Environment -
Guidance