Doppler sample volume gate position (Trackball)

Description
Adjusting
Values
Benefits
Bioeffects

Doppler sample volume length

Description
Adjusting
NOTE:
Values
Benefits
Bioeffects
5
98
-
Moves the sample volume gate on the B-Mode's Doppler Mode
cursor. The gate is positioned over a specific position within the
vessel.
To move Doppler Mode cursor position, move Trackball left or
right until positioned over the vessel.
To move sample volume gate position, move Trackball up or
down until positioned inside the vessel.
Defaults to 50% of the depth and can move continuously
throughout the field of view.
Positions the sample volume gate to sample blood flow.
Changing the sample volume gate position may change the TI
and/or MI. Observe the output display for possible effects.
Sizes the sample volume gate.
To increase/decrease the gate size, adjust SV Length on the
Touch Panel. Hold down key to continuously size gate. A black
bar appears indicating changes made to the sample volume
size, but this is not a time discontinuity.
Adjustments to the sample volume gate size are made from the
center point of the sample volume position.
Values vary by probe and application.
Sample volume gate size values are returned to the factory or
user preset value when you change the following: Probe, Exam
Category, Exam Calcs, or New Patient.
A smaller gate produces accurate sampling results because it is
more sensitive. You can also enlarge the gate to hear the
Doppler audio better or for sampling large vessels or areas.
Changing the sample volume gate size may change the TI and/
or MI. Observe the output display for possible effects.
LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1