Alarm Conditions

All alarms are low priority alarms.
Alarm
General malfunction
yellow light
and intermittent
audible alarm
Oxygen monitor
yellow light
and intermittent
audible alarm
Power failure yellow
light
and intermittent
audible alarm
Indicates
high product tank
pressure
OR
low product tank pressure
OR
high device temperature
low oxygen concentration
power failure
VisionAire Family
Action
Ensure flowmeter is open to minimum flow rate or high-
er. Ensure cannula is not kinked or obstructed. Remove
any devices connected downstream of the outlet of the
device. Ensure device has at least 12" of clearance on
all sides and intakes are not obstructed. Ensure external
gross particle intake filter is clean and not clogged. En-
sure unit is within operating temperature range. If issue
persists, contact equipment provider for service.
Contact equipment provider for service.
Ensure device is plugged into a known, working outlet.
Ensure breaker switch is pushed in. If issue persists
contact equipment provider for service.
PN MN238-C4 A | User Manual