Guidance And Manufacturer's Declaration-Electromagnetic Immunity; Immunity Test; Electromagnetic Environment - Guidance - CAIRE VisionAire 5 User Manual

Hide thumbs Also See for VisionAire 5:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Guidance and Manufacturer's Declaration—Electromagnetic Immunity
The VisionAire is intended for use in the electromagnetic environment specified below. The customer or the
user of the VisionAire should assure that it is used in such an environment.

Immunity test

IEC 60601 test level
Electromagnetic
environment –
±6 kV contact
guidance
±8 kV air
IEC 61000-4-2
±2 kV for power supply
lines
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for input/output
lines
±1 kV line(s) to line(s)
Surge
IEC 61000-4-5
±2 kV line(s) to earth
<5 % U
(>95 % dip in U
for 0,5 cycle
40 % U
(60 % dip in U
Voltage dips, short
interruptions and
for 5 cycles
voltage variations
on power supply
70 % U
input lines IEC
61000-4-11
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency
magnetic field IEC
3 A / m
61000-4-8
NOTE U
is the A.C. mains voltage prior to application of the test level.
T
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
N/A
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
<5 % U
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles
<5 % U
T
)
(>95 % dip in U
T
for 5 sec
3 A / m
VisionAire Family
Electromagnetic environment –
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
T
)
T
Mains power quality should be that of
T
)
a typical commercial or hospital envi-
T
ronment. If the user of the VisionAire
requires continued operation during
power mains interruptions, it is recom-
mended that the VisionAire is powered
T
from an uninterruptible power supply
)
T
(UPS).
T
)
T
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
PN MN238-C4 A | User Manual