PROVIDER MODE FUNCTIONS

Providers can access provider mode software functions using the Control Panel. All provider mode information is displayed on the LCD.
The software shall advance the following Eclipse 5 display mode when the "No Smoking" icon is pressed:
• Alarm Code (ALRM) - Displays most recent alarm code. Additional alarm codes
will also be displayed, if present, by continuing to press the + key.
• Pulse Sensitivity (PS)
• Hours of Operation (HRS)
• 9-volt Battery Status (9V)
• Rise Time (BOL)
• Control Board Software Revision (CS)
• Power Manager Software Revision (PM)

ALARM CODE

While in Continuous Flow Mode, pressing the "No Smoking" icon displays the current Eclipse 5 alarm code on the
LCD. This code may be used to help diagnose conditions indicated by the alert and alarm indicators. The LCD will
show "ALRM=". Refer to the "Level 1 Maintenance" section for further information.

PULSE MODE SENSITIVITY ADJUSTMENT

During patient setup, a qualified clinician may adjust the Pulse Mode sensitivity to best suit patient inspiratory effort.
The LCD will show "PS=x". PS=1 is the most sensitive setting, PS=3 is the least sensitive setting. Factory default is PS=2.

HOURS OF OPERATION

The Eclipse 5 hour meter provides valuable information on hours of operation. Providers may display the total
number of hours of operation using the Control Panel. The display counts up to "99,999"hours, then rolls over to
"00000."The LCD will show "HRS xxxxx". Refer to the "Level 1 Maintenance" section for further information.
9-VOLT BATTERY STATUS
The status of the 9-Volt Battery is provided. This is the actual voltage of the 9-Volt battery. The LCD will show "9V=x.x".

RISE TIME (BOLUS DELIVERY)

During patient set up, a qualified clinician may adjust the Rise Time (fast, medium or slow), i.e. how quickly
the set bolus volume is delivered. The LCD will display "BOL=". Factory default is BOL=FAST.

SOFTWARE VERSION

Providers may occasionally need to obtain the software version on the Eclipse 5 to perform maintenance. The
provider may display software version numbers for both the Control Board and the Power Manager software using
the Control Panel. Refer to the "Level 1 Maintenance" section for further information.
Example: CB796010: Control Board PCB: PN 7960 with Rev 1.0 software
PM796110: Power Manager PCB: PN 7961 with Rev 1.0 software
BLUETOOTH™ STATUS
OFF = off
NC = Not Connected
CNT = Connected.