Data Logger; Data Output (Printer/Computer); Calibration Validity; Date/Hour - Hach TitraLab User Manual

Ph & alkalinity content analyzer
Hide thumbs Also See for TitraLab:
Table of Contents

Advertisement

4.9.3

Data Logger

4.9.3.1
Sample numbers
4.9.4

Data Output (printer/computer)

4.9.5

Calibration validity

4.9.6

Date/Hour

4.9.7

Display contrast

4.9.8

Language

4.9.9

Theoretical Cal.

This section describes the functions of the data log.
Up to 55 results are stored.
The results can be displayed, printed or sent to a computer.
When the data log is full, the oldest result is overwritten automatically.
1. Select
SYSTEM > DATA LOGGER > DISPLAY DATA
of search filters can be used to display the results.
: displays the results for a specific date.
BY DATE
: displays all results.
CONTINUE
2. To empty the data log, select
press
to confirm.
OK
When a titration program is started, the first sample is assigned the number 1. All
subsequent samples receive sequential numbers until the titration program is closed.
When the program is launched again, the sequential numbers are reset to 1.
This function is only displayed if a printer or computer is connected and activated for data
output. Programs, settings and results can be printed or transferred. Following a titration
or calibration, the following data is printed or transferred:
Type of analysis, date and time, titration result, titrant end volume, duration of titration
and sample number.
1. Select
SYSTEM > DATA OUTPUT
2. Select
FOR PRINTER
matrix. Confirm with
Select
FOR COMPUTER
The validity of the last pH calibration can be defined in days and hours.
Entry of date and time.
Increases or decreases the display contrast.
English, Spanish, Italian, French.
When no current calibration is available, the system can revert to a theoretical calibration.
The current state of the electrode is not factored into the theoretical calibration.
SYSTEM > DATA LOGGER > EMPTY DATA-LOGGER
.
followed by the relevant printer model, such as thermal or dot
OK.
followed by either
CAPTURE
and press
to confirm. A range
OK
or
. Confirm with
TICOM
OK.
Operation
and
31