Data Logger - Hach sension ph31 User Manual

Hide thumbs Also See for sension ph31:
Table of Contents

Advertisement

Change the stirring settings
The magnetic stirrer can be turned on and the stirring speed can be
changed in the Stirring menu.
1. From the main menu use the
2. Use the
or
to select Stirring and confirm.
3. To turn on/off the stirrer push
4. When the stirrer is turned on, use the
speed in %.
Note: Use the
or
to change the stirring speed during calibration and
during a measurement.
Change the temperature units
The temperature units can be changed to Celsius or Fahrenheit.
1. From the main menu use the
2. Use the
or
to select Temperature units and confirm.
3. Use the
or
to select between Celsius or Fahrenheit and confirm.

Data logger

Display data
The Display data log contains Measurement data, Electrode report and
Calibration data. The stored data can be send to a printer or to a PC.
When the data log becomes full (400 data points), the oldest data point
is deleted when a new data point is added.
1. From the main menu use the
2. Use the
or
to select DATA LOGGER and confirm.
3. Use the
or
to select Display data and confirm.
12 English
or
to select SYSTEM. Confirm.
.
or
to change the stirring
or
to select SYSTEM. Confirm.
or
to select SYSTEM. Confirm.
4. Use the
or
to select
Option
Description
Measurement data Measurement data—stores automatically each time a
sample is measured
Electrode report
Electrode report—stores automatically the electrode
history and measurement conditions
Calibration data
Calibration data—stores automatically the current
calibration
Delete data
The entire measurement data or electrode report log can be deleted to
remove data that has already been sent to a printer or PC.
1. From the main menu use the
2. Use the
or
to select DATA LOGGER and confirm.
3. Use the
or
to select Erase and confirm.
4. Use the
or
to select Measurement data or Electrode report and
confirm. Confirm again to delete the data.
The entire log is deleted at once.
Send data to a printer or to a computer
N O T I C E
The data output (printer or PC) needs to be selected first, so that the Print menu
is available (refer to
Select the data output
Note: Refer to
Report output
on page 13 to select the report output type.
1. From the main menu use the
2. Use the
or
to select DATA LOGGER and confirm.
3. Use the
or
to select Print and confirm. Select one of the
following options and confirm with
data, Electrode data, Calibration data, Calibration report or
Instrument condit.
or
to select SYSTEM. Confirm.
on page 11).
or
to select SYSTEM. Confirm.
to print the data: Measurement

Hide quick links: