Baylis Medical RFP-100A Instructions For Use Manual page 21

Bmc radiofrequency puncture generator
Hide thumbs Also See for RFP-100A:
Table of Contents

Advertisement

Available languages

Available languages

is released, when the RF ON/OFF Button is depressed during ON mode, or when
an error is detected.
In the "DONE" mode, the RF output is OFF, the RF ON/OFF Button is not
illuminated, and the audible tone is off. The Generator passes through the DONE
mode as a transition from the ON mode back to the READY mode. The DONE
mode lasts for five seconds and then automatically changes to the READY mode.
The last measured Power, Impedance, and Timer Displays are held during the
DONE mode. The last two seconds of the DONE mode will display an error code
if termination was due to an error.
• FOOTSWITCH LABEL
"READY"
Mode
"ON"
Mode
Footswitch
Pressed
21 DMR RFP-100 3.3 V-13 16-Jan-2012_EN.doc
"DONE"
Mode
Timer elapsed
Footswitch released
ON/OFF button
pressed