5.4.

TIME

• TIME DISPLAY
The TIME Display is shown in Figure 5-1E. This display is a two-digit, green
LED numeric display. The TIME Display will read zero when the Generator is
initially powered "ON". The desired output duration expressed in seconds can be
set by the TIME ▲/▼ Buttons (Fig. 5-1D/5-1F), during the READY mode. The
count-up time duration (0-99 seconds) is shown during the ON mode, as RF
energy is delivered. If RF power delivery is stopped before the preset duration
has been reached, the actual duration of RF power delivery will be displayed for
five seconds while the generator reaches the DONE mode and transitions to the
READY mode. Once in READY mode, the prior preset time duration will be
displayed and can be adjusted as desired.
• TIME ▲/▼ BUTTONS
The TIME ▲/▼ Buttons are shown in Figure 5-1D/5-1F. Pressing the ▲/▼
buttons increase/decrease the duration that RF energy will be delivered. By
pressing the button once, you can increase the maximum duration by incremental
steps of 1 second. By continuously holding down the button, you will see the
numerical display scroll rapidly.
Note: The TIME ▲/ ▼ Buttons are NOT functional during the RF Power "ON" mode
(when RF power is being delivered).
18 DMR RFP-100 3.3 V-13 16-Jan-2012_EN.doc