EM TEST
9.
Test set up
9.1.
Test setup for Burst impulses
Coupling to power lines via coupling NX
Coupling to signal and data lines via an external
capacitive coupling clamp.
9.2.
Test level with Burst as per IEC 61000-4-4 Ed.2.
Burst generators, which comply with the specifications of IEC 61000-4-4 Ed3: 2012, have a limitation of the
maximum output voltage. Therefore, the maximum test level is limited by the number of coupling on several lines.
Please see the following limits:
Coupling
50
1 coupling any
2 couplings any
3 couplings any
Generator with
3- phase coupler
50
1 coupling any
2 couplings any
3 couplings any
4 couplings any
5 couplings any
Manual of operation
Figure 9.1
Figure 9.2
compact NX5
5500V
5500V
5000V
5000V
compact NX5
5500V
5500V
5000V
5000V
4500V
4500V
V 1.03
coupling NX