Timing Parameter Commands
Parameters
Return Value
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

:PULSe:WIDTh

Syntax
:<Handle>:SGENeral:PDATa(*)[:SOURce]:PULSe:WIDTh <Width>
:<Handle>:SGENeral:PDATa(*):TERMinal(*)[:SOURce]:PULSe:WIDTh
<Width>
:<Handle>:SGENeral:PPULse(*)[:SOURce]:PULSe:WIDTh <Width>
:<Handle>:SGENeral:PPULse(*):TERMinal(*)[:SOURce]:PULSe:WIDTh
<Width>
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*)[:SOURce]:PULSe:WIDTh
<Width>
<Width>
Pulse width in seconds (<NRf>).
Sets the width or duration of the pulse.
Valid range 0 ... PERiod
(on connector level: 0 ... PERiod / Connector:MUX).
Example
:_test:SGEN:PDAT1:PULS:WIDT 15e-9
:_test:SGEN:PDAT1:TERM1:SOUR:PULS:WIDT 10e-9
:_test:SGEN:PPUL1:PULS:WIDT 15e-9
:_test:SGEN:PPUL1:TERM1:SOUR:PULS:WIDT 10e-9
:_test:MOD2:CONN3:PULS:WIDT 20e-9
:PULSe:WIDTh?
Syntax
:<Handle>:SGENeral:PDATa(*):[:SOURce]:PULSe:WIDTh?
:<Handle>:SGENeral:PDATa(*):TERMinal(*)[:SOURce]:PULSe:WIDTh?
:<Handle>:SGENeral:PPULse(*):[:SOURce]:PULSe:WIDTh?
:<Handle>:SGENeral:PPULse(*):TERMinal(*)[:SOURce]:PULSe:WIDTh?
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*)[:SOURce]:PULSe:WIDTh?
Returns the current pulse width value.
Example
:_test:SGEN:PDAT1:PULS:WIDT?
:_test:SGEN:PDAT1:TERM1:SOUR:PULS:WIDT?
:_test:SGEN:PPUL1:PULS:WIDT?
:_test:SGEN:PPUL1:TERM1:SOUR:PULS:WIDT?
:_test:MODULE2:CONNECTOR3:PULSE:WIDTH?
might return:
1.000000E-008