Level Parameter Commands
Commands in DSR Application Subsystems

:VOLTage[:LEVel][:IMMediate]:LOW

Syntax
:<Handle>:SGENeral:PDATa(*)[:SOURce]:VOLTage[:LEVel][:IMMediate]:LOW
<LowLevel>
:<Handle>:SGENeral:PDATa(*):TERMinal(*)[:SOURce]:VOLTage[:LEVel]
[:IMMediate]:LOW <LowLevel>
:<Handle>:SGENeral:PPULse(*)[:SOURce]:VOLTage[:LEVel][:IMMediate]:LOW
<LowLevel>
:<Handle>:SGENeral:PPULse(*):TERMinal(*)[:SOURce]:VOLTage[:LEVel]
[:IMMediate]:LOW <LowLevel>
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*)[:SOURce]:VOLTage[:LEVel]
[:IMMediate]:LOW <LowLevel>
Parameters
<LowLevel>
Low Level value in Volts (<NRf>).
Sets the low level of the time varying signal.
Valid range depends on the frontend used:
E4838A: –2.2 to 4.35 V
E4843A: –2.20 to 2.95 V
These ranges are valid for a termination voltage of 0V (50 Ohm). With
other termination voltages other values can be reached (for example,
PECL Levels).
Example
:_test:SGEN:PDAT1:VOLT:LOW -1
:_test:SGEN:PDAT1:TERM1:SOUR:VOLT:LEV:IMM:LOW -1
:_test:SGEN:PPUL1:VOLT:LOW -1
:_test:SGEN:PPUL1:TERM1:SOUR:VOLT:LEV:IMM:LOW 0.1
:_test:MOD2:CONN4:VOLT:LOW 0.3
227
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006