Handle>:Sgeneral:pdata:dmux:stage - Agilent Technologies 81250 System Programming Manual And Scpi Reference

Parallel bit error ratio tester
Hide thumbs Also See for 81250:
Table of Contents

Advertisement

Commands in DSR Application Subsystems
Return Value
Description
Parameters
222
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:<Handle>:SGENeral:GLOBal:
SYNChronization:RCH?
Syntax
:<Handle>:SGENeral:GLOBal:SYNChronizat:RCHeck?
Returns the current rewire verify mode (on or off).
If verify rewiring is switched on, the system performs additional
inspections and complains about possible errors.
If verify rewiring is switched off, the system tries to ignore noncritical
problems to carry out a measurement.
Example
:_test:SGEN:GLOB:SYNC:RCH?
might return:
ON or OFF
:<Handle>:SGENeral:PDATa:DMUX:STAGe
Syntax
:<Handle>:SGENeral:PDATa*:DMUX:STAGe*[:VALUE]
Specifies the number of stages of the DeMUX module. To set this value,
you need to know the architecture of the demultiplexer.
*. Number of stages of the DeMUX module.
Example
For a demultiplexer with the following architecture:
Stage
1
:_test:SGEN:PDATA:DMUX:STAG 3
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
DeMUX Rewiring Parameter Commands
Stage
Stage
2
3

Advertisement

Table of Contents
loading