:SYSTem Subsystem
Description
Parameters
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:<Handle>:SYSTem:TEST:MODule
Syntax
:<Handle>:SYSTem:TEST:MODule DETailed | SHORt [,CGR#, MOD#]
A module selftest is started by the firmware on every module. The
results of this test are stored in the message queue. The message queue
can be read by with the command :SYST:MQU?.
The format of the messages can be specified by following keywords:
<DETailed>
<SHORt>
Generates a less detailed message.
[,CGR#, MOD#]
parameters. The first parameter (CGR#) is the clock group the module
belongs to, the second parameter (MOD#) is the number of the module
itself.
Example
For a SHORt message:
(1,3,1,4,'E4861A','Passed','E4861A Ver: 1.92 E4863 E4862 ','No
error')
For a DETailed message:
= (1,1,1,2,'E4805B','Failed','< AGILENT TECHNOLOGIES, E4805B, 01,
<>, "", Module Software Version = 1.81, Boot Loader Version = 1.00
>','-330, Self-test failed: Frequency generation PLL test failed:
high limit not reached!-330, Self-test failed: Frequency generation
PLL test failed: high limit not reached!')
A DETailed message for an E4861A could appear as:
(1,4,1,7,'E4861A','Failed','< AGILENT TECHNOLOGIES, E4861A, 00, <>,
"", Module Software Version = 0.32, Boot Loader Version = 0.01,
<E4863, 01, <>, "">, <E4862, 01, <>, ""> >','-330, Self-test
failed: Voltage of +5VRef is out of valid range')
Commands in DSR Application Subsystems
Generates a more detailed message.
A single module can be specified by the optional
81