GRATING SPECTROPHOTOMETER OPERATING MANUAL
Select "Calibration" to enter white and black calibration interface as
shown in Figure 5.
It will show if the calibration is valid or not, and the remaining time if it
is valid.
Click
to continue and enter "White Calibration" as shown in Figure
2. According to warning, confirm the serial number of white calibration
plate correspond to the instrument and correct aperture setting, then put
the measuring aperture on white board and press "Testing" button for
white calibration, or click
After white calibration, it will prompt you to black calibration as shown
in Figure 6. According to warning, put the measuring aperture on black
board and press "Testing" button for black calibration, or click
cancel and quit the calibration.
Figure 5 White and Black Calibration
to cancel and quit the calibration.