Identifying Battery Capacity Using the Cadex SMart Two+ Charger
Identifying Battery Status Using the Battery Status Menu
4-14
Maintenance: Lithium-Ion Battery
When you select the Target Capacity switch on the charger, the
charger compares the battery's performance to a 60%, 70%, or 80% target
capacity set on the battery charger. If the battery fails to meet the target
performance, the battery charger will prompt you to condition the
battery. If after the condition cycle has been completed, the battery does
NOT meet the target performance, then the battery charger will
illuminate a "fail" light.
This menu option opens a menu and information window that provides
current battery status information.
1. Select MORE MENUS from the Main Menu.
2. Select BATTERY STATUS to display the Battery Status Menu and
information window.
BATTERY STATUS Menu and Information Window
If a battery is not present, a NO BATT message is displayed in
this column.
If the battery is NO COMM (communication with this battery
has failed), "unknown" is displayed for all rows except the SLOT
STATUS row.
Dash 3000/4000 Patient Monitor
2000966-339
Revision A