Analog Out Buzz

Completion

6-14
Configuration: Boot Code Selections
Select Analog Out Buzz to identify an ECG or an invasive BP "leads off"
condition via a supported analog out device. The default is 0 volts.
1. Activate the Boot Code.
a. Hold down NBP GO/STOP and ZERO ALL.
b. Press and release the Trim Knob control.
c.
Keep holding NBP GO/STOP and ZERO ALL until the Boot
Code information appears on the display.
2. Select the SERVICE MENU.
3. Select the SET CONFIGURATION menu option.
4. Select the Analog Out Buzz menu option.
5. Select 60Hz sine wave to display a 60 Hz sine wave, or select 0 volts
to diaplay a flat line.
6. Select Exit and wait for the monitor to display the message, "Writing
settings to EEPROM..."
7. Manually reboot the monitor to implement this change.
The monitor is now ready for normal operation. At this time, perform the
"Checkout Procedure" found in Chapter 4, Maintenance.
Dash 3000/4000 Patient Monitor
2000966-339
Revision A