Micropilot FMR51, FMR52
Endress+Hauser
Output
The Micropilot is commissioned by entering an empty distance "E" (=zero), a full distance "F" (=span)
and application parameters which automatically adapt the instrument to the process conditions. For
models with a current output, the factory adjustment for zero point "E" and span "F" is 4 mA and 20
mA. For digital outputs and the display module, the factory adjustment for zero point "E" and span "F"
is 0 % and 100 %.
A linearization with max. 32 points, based on a table entered either manually or semi-automatically,
can be activated locally or remotely. This function provides a measurement in engineering units and
a linear output signal for spheres, horizontal cylindrical tanks and vessels with conical outlet.
Life cycle of the product
Engineering
• Universal measuring principle
• Measurement unaffected by medium properties
• Hardware and software developed according to SIL IEC 61508
Procurement
• Endress+Hauser being the world market leader in level measurement guarantees asset protection
• Worldwide support and service
Installation
• Special tools are not required
• Reverse polarity protection
• Modern, detachable terminals
• Main electronics protected by a separate connection compartment
Commissioning
• Fast, menu-guided commissioning in only a few steps on site or from the control room
• Plain text display in national languages reduces the risk of error or confusion
• Direct local access of all parameters
• Short instruction manual at the device
Operation
• Multi-echo tracking: Reliable measurement through self-learning echo-search algorithms taking
into account the short-term and long-term history in order to check the found echoes for
plausibility and to suppress interference echoes.
• Diagnostics in accordance with NAMUR NE107
Maintenance
• HistoROM: Data backup for instrument settings and measured values
• Exact instrument and process diagnosis to assist fast decisions with clear details concerning
remedies
• Intuitive, menu-guided operating concept in national languages saves costs for training,
maintenance and operation
• Cover of the electronics compartment can be opened in hazardous areas
Retirement
• Order code translation for subsequent models
• RoHS-conforming (Restriction of certain Hazardous Substances), unleaded soldering of electronic
components
• Environmentally sound recycling concept
7
Need help?
Do you have a question about the Micropilot FMR51 and is the answer not in the manual?
Questions and answers