Endress+Hauser Micropilot FMR51 Technical Information page 23

Free space radar
Hide thumbs Also See for Micropilot FMR51:
Table of Contents

Advertisement

Micropilot FMR51, FMR52
Endress+Hauser
4-wire: 4-20mA HART (10.4 to 48 V
13
12
11
≥250 Ω
1
2 3
 5
Terminal assignment 4-wire; 4-20mA HART (10.4 to 48 V
1
Evaluation unit, e.g. PLC
2
HART communication resistor (≥ 250 Ω): Observe maximum load
3
Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4
Analog display device: Observe maximum load
5
Signal cable including screening (if required), observe cable specification
6
Protective connection; do not disconnect!
7
Protective earth, observe cable specification
8
4...20mA HART (active): Terminals 3 and 4
9
Supply voltage: Terminals 1 and 2
10
Supply voltage: Observe terminal voltage, observe cable specification
11
Terminal for potential equalization
12
Cable entry for signal line
13
Cable entry for power supply
)
DC
A
10
9
+
-
4...20 mA
4
5
)
DC
8
7
6
A0011340
23

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Micropilot FMR51 and is the answer not in the manual?

This manual is also suitable for:

Micropilot fmr52Micropilot fmr53Micropilot fmr54

Table of Contents