PCE Instruments PCE-ITM 20 User Manual page 25

Isolation meter
Table of Contents

Advertisement

Multimeter
Measuring parameters
DC
AC
AC + DC
50 .... 1000 Hz
The specified accuracy for alternating voltage refers to 5 ... 100% of the measuring
range
Direct current
Alternating current
AC + DC
50 .... 1000 Hz
The specified accuracy for alternating current refers to 5 ... 100% of the measuring
range
Measuring range
50 mV
500 mV
5 V
50 V
500 V
1000 V
50 mV
500 mV
5 V
50 V
500 V
1000 V
Measuring range
500 µA
5000 µA
50 mA
500 mA
10 A
500 µA
5000 µA
50 mA
500 mA
10 A
© PCE Instruments
21
Resolution
Accuracy
0.001 mV
± (0.06 %+ 20 digits)
0.01 mV
± (0.06 %+ 6 digits)
0.0001 V
± (0.06 %+ 4 digits)
0.001 V
± (0.06 %+ 4 digits)
0.01 V
± (0.06 %+ 4 digits)
0.1 V
± (0.1 %+ 3 digits)
0.001 mV
± (1 % + 50 digits)
0.01 mV
± (1 % + 50 digits)
0.0001 V
± (1 % + 30 digits)
0.001 V
± (1 % + 30 digits)
0.01 V
± (1 % + 30 digits)
0.1 V
± (1 % + 30 digits)
Resolution
Accuracy
0.01 µA
± (0.5 % + 3 digits)
0.1 µA
± (0.5 % + 3 digits)
0.001 mA
± (0.5 % + 3 digits)
0.01 A
± (0.5 % + 3 digits)
0.001 A
± (0.5 % + 3 digits)
0.01 µA
± (1% + 30 digits)
0.1 µA
± (1% + 30 digits)
0.001 mA
± (1% + 30 digits)
0.01 A
± (1% + 30 digits)
0.001 A
± (1% + 30 digits)