4.2
Replacing the Fuses
To avoid electric shock, disconnect the test leads from any source of voltage
before removing the meter cover.
• Disconnect the test leads from the meter.
• Remove the protective rubber holster.
• Remove the battery cover (two "B" screws) and the battery.
• Remove the six "A" screws securing the rear cover.
• Gently remove the old fuse and install the new fuse into the holder.
• Always use a fuse of the proper size and value (0.5A/1000V fast blow for the 500mA range,
10A/1000V fast blow for the 20A range
• Replace and secure the rear cover, battery and battery cover.
To avoid electric shock, do not operate your meter until the fuse cover is in
place and fastened securely.
5
Technical specifications
Test voltage / test
current
50V 1-mA @ 50 kΩ
125V / 1-mA @ 125 kΩ
250V / 1-mA @ 250 kΩ
500V / 1-mA @ 500 kΩ
1000V / 1-mA @ 1000
kΩ
Short circuit current
Measuring
Resolution
range
0.01 ... 5 MΩ
0.0001 MΩ
5 ... 50 MΩ
0.001 MΩ
50 ... 500 MΩ
0.01 MΩ
500 ... 1000 MΩ
0.1 MΩ
0.01 ... 5 MΩ
0.001 MΩ
5 .... 50 MΩ
0.001 MΩ
50 ... 500 MΩ
0.01 MΩ
500 .... 5000 MΩ
0.1 MΩ
0.01 ... 5 MΩ
0.0001 MΩ
5 ... 50 MΩ
0.001 MΩ
50 ... 500 MΩ
0.01 MΩ
500 ... 5000 MΩ
0.1 MΩ
0.01 ... 5 MΩ
0.0001 MΩ
5 ... 50 MΩ
0.001 MΩ
50 ... 500 MΩ
0.01 MΩ
500 ... 5000 MΩ
0.1 MΩ
0.01 ... 5 MΩ
0.0001 MΩ
5 ... 50 MΩ
0.001 MΩ
50 ... 500 MΩ
0.01 MΩ
500 ... 5000 MΩ
0.1 MΩ
<1.5-mA
© PCE Instruments
20
Accuracy
± (4 % + 20 digits)
± (4 % + 20 digits)
± (4 % + 20 digits)
not specified
± (2 % + 20 digits)
± (2 % + 20 digits)
± (3 % + 20 digits)
± (4 % + 20 digits)
± (2 % + 20 digits)
± (2 % + 20 digits)
± (3 % + 20 digits)
± (4 % + 20 digits)
± (2 % + 20 digits)
± (2 % + 20 digits)
± (3 % + 20 digits)
± (4 % + 20 digits)
± (2 % + 20 digits)
± (2 % + 20 digits)
± (3 % + 20 digits)
± (4 % + 20 digits)