Replacing The Fuses; Technical Specifications - PCE Instruments PCE-ITM 20 User Manual

Isolation meter
Table of Contents

Advertisement

4.2

Replacing the Fuses

To avoid electric shock, disconnect the test leads from any source of voltage
before removing the meter cover.
• Disconnect the test leads from the meter.
• Remove the protective rubber holster.
• Remove the battery cover (two "B" screws) and the battery.
• Remove the six "A" screws securing the rear cover.
• Gently remove the old fuse and install the new fuse into the holder.
• Always use a fuse of the proper size and value (0.5A/1000V fast blow for the 500mA range,
10A/1000V fast blow for the 20A range
• Replace and secure the rear cover, battery and battery cover.
To avoid electric shock, do not operate your meter until the fuse cover is in
place and fastened securely.
5

Technical specifications

Test voltage / test
current
50V 1-mA @ 50 kΩ
125V / 1-mA @ 125 kΩ
250V / 1-mA @ 250 kΩ
500V / 1-mA @ 500 kΩ
1000V / 1-mA @ 1000
Short circuit current
Measuring
Resolution
range
0.01 ... 5 MΩ
0.0001 MΩ
5 ... 50 MΩ
0.001 MΩ
50 ... 500 MΩ
0.01 MΩ
500 ... 1000 MΩ
0.1 MΩ
0.01 ... 5 MΩ
0.001 MΩ
5 .... 50 MΩ
0.001 MΩ
50 ... 500 MΩ
0.01 MΩ
500 .... 5000 MΩ
0.1 MΩ
0.01 ... 5 MΩ
0.0001 MΩ
5 ... 50 MΩ
0.001 MΩ
50 ... 500 MΩ
0.01 MΩ
500 ... 5000 MΩ
0.1 MΩ
0.01 ... 5 MΩ
0.0001 MΩ
5 ... 50 MΩ
0.001 MΩ
50 ... 500 MΩ
0.01 MΩ
500 ... 5000 MΩ
0.1 MΩ
0.01 ... 5 MΩ
0.0001 MΩ
5 ... 50 MΩ
0.001 MΩ
50 ... 500 MΩ
0.01 MΩ
500 ... 5000 MΩ
0.1 MΩ
<1.5-mA
© PCE Instruments
20
Accuracy
± (4 % + 20 digits)
± (4 % + 20 digits)
± (4 % + 20 digits)
not specified
± (2 % + 20 digits)
± (2 % + 20 digits)
± (3 % + 20 digits)
± (4 % + 20 digits)
± (2 % + 20 digits)
± (2 % + 20 digits)
± (3 % + 20 digits)
± (4 % + 20 digits)
± (2 % + 20 digits)
± (2 % + 20 digits)
± (3 % + 20 digits)
± (4 % + 20 digits)
± (2 % + 20 digits)
± (2 % + 20 digits)
± (3 % + 20 digits)
± (4 % + 20 digits)

Hide quick links: