Inficon IO1000 Translation Of The Original Operating Instructions
Inficon IO1000 Translation Of The Original Operating Instructions

Inficon IO1000 Translation Of The Original Operating Instructions

I/o module

Advertisement

T R A N S L A T I O N
O P E R A T I N G
IO1000
Type designation
Product description
I/O module
Catalog no.
560-310
from software
I/O module 1.11
version
jiqc10en1-c (1408)
Document no.
O F
T H E
I N S T R U C T I O N S
O R I G I N A L

Advertisement

Table of Contents
loading

Summary of Contents for Inficon IO1000

  • Page 1 T H E O R I G I N A L O P E R A T I N G I N S T R U C T I O N S IO1000 Type designation Product description I/O module Catalog no.
  • Page 2 This document applies to the software version stated on the cover page. Documents for other software versions are available from our sales department. Reprint, translation and duplication need to be approved in writing by INFICON GmbH.
  • Page 3: Table Of Contents

    Content About this manual ......4 Target groups ..........4 Other applicable documents .
  • Page 4: About This Manual

    About this manual Target groups This installation manual is intended for the owner and for technically qualified personnel with experience in leak detection technology and integration of leak detection devices in leak detection systems. In addition, the installation and use of the unit require knowledge of electronic interfaces.
  • Page 5: Text Markings

    1.3.2 Text markings Marking Meaning Requirement for execution of an action  Tool or aid for an action  ► Instruction 1, 2, 3, ... Several instructions in a fixed order  Result of an action About this manual...
  • Page 6: Safety

    Safety Intended use The I/O module is a device interface between a leak detector and an external controller. ► Install, operate and service the unit only in compliance with these instructions. ► Comply with the limits of application (see chapter 4.3, page 14).
  • Page 7: Shipment, Transport, Storage

    Shipment, transport, storage Scope of delivery Article Quantity I/O module Operating instructions ► Please check the scope of delivery of the product for completeness after receipt. Transport Damage due to unsuitable packaging material Transport in unsuitable packaging material can damage the unit. ►...
  • Page 8: Description

    Description Construction of the I/O module Fig. 1 Front view PLC IN, digital inputs DIP switch cover and fuse RS232 LD, connection of leak detector Status LED ANALOG IN, analog inputs ANALOG OUT, analog outputs RS485 PLC OUT, digital outputs 24 V OUT, 24 volt output PLC IN a Digital inputs...
  • Page 9 Fig. 2 Sample wiring, digital inputs with PLC. Left: Driver outputs. Right: Potential-free contacts PLC OUT e Digital outputs Galvanic isolation (max. 60 V DC, 25 V VAC against GND) Max. permissible load per output: U = 30 V, I = 0.75 A Fuses for digital outputs 1 ...
  • Page 10 Fig. 3 Sample wiring, digital outputs ANALOG IN h Analog input (input voltage range 0 V to 10.8 V) Pin assignment: 24V supply (output) GND to 24V supply Analog input (0 V to 10.8 V) GND to analog input ANALOG OUT d Analog outputs (e.g.
  • Page 11 Pin assignment: Analog output 1: ANALOG-OUT 1 Analog output 2: ANALOG-OUT 2 GND to analog output The output pins can be freely configured. RS232 b Connection for RS-232 Galvanic isolation (max. 60 V DC, 25 V VAC against GND) Pin assignment: Pin Name A normal RS-232 cable must be used for the connection (1:1 connection, RxD and TxD not crossed, no zero-modem cable).
  • Page 12 RS485 i Connection for RS-485 Fig. 5 RS-485 bus terminator A: Active bus terminator B: Without bus terminator C: Passive bus terminator Galvanic isolation (max. 60 V DC, 25 V VAC against GND) Pin assignment: Pin Name Connect pull-up resistor (562 Ω against +5 V) with D+ if necessary Connect pull-down (562 Ω...
  • Page 13 FUSE and DIP switch S1, S2 f Fuses for digital outputs and DIP switches (under the cover) Fuses for digital outputs 1 ... 4 and 5 ... 8: 2 x 0.75 A (Schurter: 7010.9800.xx) Contact DIP switch S1 LDS3000 HLD6000 4 3 2 1 Factory setting (default value of the interface 0 0 0 0...
  • Page 14: Function

    Function The I/O module is a device interface between the leak detector and an external controller. The I/O module is equipped with – one RS-232 connection – one RS-485 connection – one analog input – ten digital inputs – two analog outputs –...
  • Page 15: Installation And Removal

    Installation and removal Mount I/O module on DIN-TS35 top hat rail Fig. 6 Mounting of I/O module 1 Hook unit on top hat rail at bottom. 2 Press unit onto top hat rail at top. 5.1.1 Establish connections Connecting I/O module with leak detector The I/O module communicates via data cable with the leak detector and is supplied with voltage by the data cable.
  • Page 16: Remove The I/O Module From The Din-Ts35 Top Hat Rail

    Remove the I/O module from the DIN-TS35 top hat rail Fig. 7 Removing the I/O module 1 Use the flat-tip screwdriver to pull out the locking device. 2 Pull the device off of the top hat rail. Installation and removal...
  • Page 17: Disposal

    Disposal The unit can be disposed of by the owner or sent to INFICON. The unit is made of materials that can be reused. You should use this option to avoid waste and save the environment. ► For disposal, always comply with local and regional environmental and safety regulations.
  • Page 18 INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, Germany UNITED STATES TAIWAN JAPAN KOREA SINGAPORE GERMANY FRANCE UNITED KINGDOM HONG KONG Vi sit o ur we bsi t e fo r con t act i nfo rm ati on an d o t her sal es o ffice s wo rl dw id e. w w w . i n f i c o n . c o m...