Advertisement

Translation of the operating instructions
IO1000
I/O-Modul
Catalog No.
560-310
From software version
I/O-Modul 1.14
jiqc10en1-05-(1901)

Advertisement

Table of Contents
loading

Summary of Contents for Inficon IO1000

  • Page 1 Translation of the operating instructions IO1000 I/O-Modul Catalog No. 560-310 From software version jiqc10en1-05-(1901) I/O-Modul 1.14...
  • Page 2 INFICON GmbH Bonner Strasse 498 50968 Cologne, Germany...
  • Page 3: Table Of Contents

    5 Installation and removal.......................... 17 5.1 Mount device on DIN-TS35 top hat rail.................... 17 5.1.1 Establish connections .........................  17 5.2 Remove the I/O module from the DIN-TS35 top hat rail................. 18 6 Disposal................................ 19 7 CE Declaration of Conformity ......................... 20 8 RoHS ................................ 21 IO1000-operating-instructions-jiqc10en1-05-(1901)
  • Page 4: About These Instructions

    Imminent hazard resulting in death or serious injuries WARNING Hazardous situation resulting in potential death or serious injuries CAUTION Hazardous situation resulting in minor injuries NOTICE Hazardous situation resulting in damage to property or the environment 4 / 24 IO1000-operating-instructions-jiqc10en1-05-(1901)
  • Page 5: Safety

    2.3 Operator requirements ► Read, observe and follow the information in these instructions and the working instructions created by the owner, especially the safety instructions and warnings. ► Perform all work based on the complete instructions. IO1000-operating-instructions-jiqc10en1-05-(1901) 5 / 24...
  • Page 6: Scope Of Delivery

    ► Please check the scope of delivery of the product for completeness after receipt. Transport NOTICE Damage due to unsuitable packaging material Transport in unsuitable packaging material can damage the device. ► Transport the device only in the original packaging material. ► Keep original packaging material. 6 / 24 IO1000-operating-instructions-jiqc10en1-05-(1901)
  • Page 7: Description

    PLC IN, digital inputs DIP switch cover and fuse RS232 LD, connection of leak detector Status LED ANALOG IN, analog inputs ANALOG OUT, analog outputs RS485 PLC OUT, digital outputs 24 V OUT, 24 volt output IO1000-operating-instructions-jiqc10en1-05-(1901) 7 / 24...
  • Page 8 A signal on these digital inputs must have a minimum length of 100 ms, so that it can be safely evaluated. Pin assignment: The input pins PLC-IN 1 to PLC-IN 10 can be configured freely, see example: Fig. 2: Sample wiring, digital inputs with PLC. Left: Driver outputs, right: Potential-free contacts 8 / 24 IO1000-operating-instructions-jiqc10en1-05-(1901)
  • Page 9 U = 30 V, I = 0.75 A Fuses for digital outputs 1 ... 4 and 5 ... 8: 2 x 0.75 A Pin assignment: The output pins PLC_OUT_1 to PLC_OUT_8 can be configured freely, see example: Fig. 3: Sample wiring, digital outputs IO1000-operating-instructions-jiqc10en1-05-(1901) 9 / 24...
  • Page 10 Pin assignment: Analog output 1: ANALOG-OUT 1 Analog output 2: ANALOG-OUT 2 GND to analog output The output pins can be freely configured. Specific information can be found in the operating instructions of the respective leak detector. 10 / 24 IO1000-operating-instructions-jiqc10en1-05-(1901)
  • Page 11 TxD not crossed, no zero-modem cable). ► Deactivate RS-232 hardware handshake in RS-232 control program. If the hardware handshake cannot be deactivated, the RS-232 can be used as follows: Fig. 4: Connection with RS-232 cable (in case hardware handshake cannot be deactivated) IO1000-operating-instructions-jiqc10en1-05-(1901) 11 / 24...
  • Page 12 The bus terminator integrated in the I/O module (120 Ω) between D+ and D can be disabled via DIP switches S2-4. The BUS address is 1. BUS operation with more than two subscribers is not possible. 12 / 24 IO1000-operating-instructions-jiqc10en1-05-(1901)
  • Page 13 1 = ON, 0 = OFF DIP switch S2 Contact Activate boot mode for software update Disable bus terminator 120 Ω for RS-485 1 = ON, 0 = OFF, + = switching from OFF to ON during operation, X = random IO1000-operating-instructions-jiqc10en1-05-(1901) 13 / 24...
  • Page 14 The I/O module is supplied with voltage by the leak detector and requires no separate power supply. The 24V output is not used for voltage supply to the I/O module. The 24V output of the I/O module can be used as an active signal for the PLC inputs and outputs. 14 / 24 IO1000-operating-instructions-jiqc10en1-05-(1901)
  • Page 15: Function

    • one RS-232 connection • one RS-485 connection • one analog input • ten digital inputs • two analog outputs • eight digital outputs The actual function is determined by the software of the connected leak detector. IO1000-operating-instructions-jiqc10en1-05-(1901) 15 / 24...
  • Page 16: Technical Data

    2000 m Max. relative humidity above 40 °C Max. relative humidity from 31 °C to 40 80% to 50% (linear abfallend) °C Max. relative humidity to 40 °C Storage temperature -20 °C - 60 °C Pollution degree 16 / 24 IO1000-operating-instructions-jiqc10en1-05-(1901)
  • Page 17: Installation And Removal

    Connect I/O module via desired interfaces with external controller: – RS232 (RS-232 interface) – RS485 (RS-485 interface) – Analog In (analog input) – Analog Out (analog outputs) – PLC In (digital inputs) – PLC Out (digital outputs) IO1000-operating-instructions-jiqc10en1-05-(1901) 17 / 24...
  • Page 18: Remove The I/O Module From The Din-Ts35 Top Hat Rail

    INFICON 5.2 Remove the I/O module from the DIN-TS35 top hat rail Fig. 7: Removing the I/O module Use the flat-tip screwdriver to pull out the locking device. Pull the device off of the top hat rail. 18 / 24 IO1000-operating-instructions-jiqc10en1-05-(1901)
  • Page 19: Disposal

    The device consists of materials that can be recycled. This option should be exercised to prevent waste and also to protect the environment. ► For disposal, always comply with local and regional environmental and safety regulations. IO1000-operating-instructions-jiqc10en1-05-(1901) 19 / 24...
  • Page 20: Ce Declaration Of Conformity

    7 | CE Declaration of Conformity INFICON 7 CE Declaration of Conformity 20 / 24 IO1000-operating-instructions-jiqc10en1-05-(1901)
  • Page 21: Rohs

    INFICON RoHS | 8 8 RoHS IO1000-operating-instructions-jiqc10en1-05-(1901) 21 / 24...
  • Page 22 8 | RoHS INFICON 22 / 24 IO1000-operating-instructions-jiqc10en1-05-(1901)

Table of Contents