Alarm Setup Of Each Parameter - EDAN M3A User Manual

Vital signs monitor
Hide thumbs Also See for M3A:
Table of Contents

Advertisement

M3A Vital Signs Monitor User Manual
Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of
alarm volume to a low level or off during patient monitoring may result in patient danger.
Remember that the most reliable method of patient monitoring combines close personal
surveillance with correct operation of monitoring equipment.
NOTE:
1 The monitor does not have alarm condition delay or alarm signal generation delay.
2 When alarms of different levels occur at the same time, the monitor prompts one of the
highest levels.
3 If the monitor is powered off and then turned on, the alarm setup can resume to the
setup which is set before the power-off.
5.1.3 Alarm Setup
Set Alarm Setup in Main Menu to open the submenu as shown below. The user can turn ON or
OFF the alarm for each parameter, and set the upper alarm limit and lower alarm limit for each
parameter by High or Low.

Alarm setup of each parameter

In the Alarm Setup menu, set the alarm limit for each parameter: SYS, DIA, MAP, SpO
For example: Method to set systolic blood pressure alarm limit for SYS alarm:
Step 1: Set SYS to ON;
Step 2: Set High (higher limit of SYS alarm) and Low (lower limit of SYS alarm).
The user can press the UP/DOWN and OK button to set the menu.
The method for setting the alarm limits of other parameters is the same as SYS alarm.
WARNING
Figure 5-1 Alarm Setup
- 37 -
Alarm
, PR.
2