M3A Vital Signs Monitor User Manual
Technical alarms: (display in the area below the NIBP value):
Message
NIBP
COMM
NIBP module failure or
STOP
communication failure.
NIBP
The hardware pressure is
ILLEGAL
too high
RESET
NIBP
The NIBP module has
MODULE ERR
failure.
Cuff
LOOSE CUFF
wrapped or no cuff exists.
Cuff, hose or connector is
AIR LEAK
damaged.
AIR
The airway of NIBP has
PREESURE
failure.
ERR
NIBP SIGNAL
Cuff is too loose or patient
TOO WEAK
pulse is too weak.
Because of arm motion,
NIBP
NOISE
signal noise is too large or
SIGNAL
pulse rate is not regular.
Pressure has exceeded the
OVER
specified
PRESSURE
limit.
NIBP SIGNAL
Excessive motion.
SATURATED
CUFF
TYPE
Cuff type does not comply
ERR
with the patient type.
Cause
is
no
properly
upper
safety
- 64 -
Alarm Level
Stop using measuring function
of
High
biomedical
manufacturer's service staff.
Measure
High
persists, stop using measuring
function of NIBP module and
notify biomedical engineer or
High
manufacturer's service staff.
Low
Properly wrap the cuff.
Low
Check and replace the leaking
parts,
biomedical
Low
manufacturer's service staff.
Use other method to measure
Low
blood pressure.
Make sure that the patient
Low
under
motionless.
Measure
persists, stop using measuring
Low
function of NIBP module and
notify biomedical engineer or
manufacturer's service staff.
Low
Stop the patient from moving.
Low
Select appropriate cuff type
NIBP Monitoring (Optional)
What to do
NIBP
module,
engineer
again,
if
failure
if
required,
engineer
monitoring
again,
if
failure
notify
or
notify
or
is