Alarm Setup - EDAN M8 Series User Manual

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Alarm Sound
The high/medium/low-level alarms are indicated by the system in following different audio ways:
Alarm level
High
Medium
Low
The sound pressure range for audible alarm signals is from 45 dB to 85 dB.
1. Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level or off during patient monitoring may result
in patient danger. Remember that the most reliable method of patient monitoring
combines close personal surveillance with correct operation of monitoring equipment.
2. Ensure the volume is properly set up. When the sound pressure of audible alarm is
below or equivalent to the ambient noise, it may be difficult for the operator to
distinguish the audio alarm.
NOTE:
1
When alarms of different levels occur at the same time, the monitor prompts the one
of the highest level.
2
If the monitor is powered off and then turned on, the alarm setup can resume to the
setup which is set before the power-off.

6.1.3 Alarm Setup

Alarm setup in the ALARM SETUP menu
Press the ALARM SETUP button in the SYSTEM SETUP menu to call up ALARM SETUP
menu (default menu) as shown below. In the ALM SEL item, the user may set up the
information about common alarm setup and the alarm setup of each parameter.
 COMMON ALM SETUP
Select COMMON ALM SETUP option in ALM SEL item. This operation may call up the
dialog box as the default one.
ALARM VOL: set the alarm volume by this item, the valid range is from 1 to 10.
ALM REC TIME: this item can be set to 8 S, 16 S and 32 S.
ALM RESET: selecting this item can reset alarm.
 Alarm setup of each parameter
Mode is "DO-DO-DO------DO-DO, DO-DO-DO------DO-DO", which
is triggered once every 10 seconds.
Mode is "DO-DO-DO", which is triggered once every 25 seconds.
Mode is "DO-", which is triggered once every 30 seconds.