Product Specifications; Electrical; Battery; Environmental - CareFusion Alaris Gateway Workstation Directions For Use Manual

Table of Contents

Advertisement

Alaris™ Gateway Workstation v1.2

Product Specifications

Product Specifications

Electrical

Protection Against Electrical Shock:
Class I
Supply Voltage:
115-230V, ~50-60Hz
Rating:
460VA (Maximum)
Fuses:
2 x 4AH Time-Lag T
Consult Qualified Service Personnel for replacement of fuses .
AC Outlets:
MDI Tile: 115-230V, ~50-60Hz, 60VA
Protection against fluid ingress:
IPX1 - Protected against vertically falling drops of water .

Battery

Type:
Nickel Metal Hydride
Charge time:
With no pumps attached: Up to 8 hours to 95% charge time based on battery life cycle .
With pumps attached: Up to 16 hours to 95% charge time based on battery life cycle and an 11 pump load .
Operating time:
20 minutes

Environmental

Operating
Transport and Storage
Temperature:
+5°C - +40°C
-20°C - +50°C
Humidity:
20% - 90%
15% - 95%
Atmospheric Pressure:
700 - 1060hPa
500 - 1060hPa

Classification

Continuous Operation

Regulatory Compliance

Complies with IEC/EN60601-1, IEC/EN60601-1-1, IEC/EN60601-1-2 .

Potential Equalisation Conductor

The function of the Potential Equalisation Connector (Conductor) is to provide a direct connection between the Workstation and the
potential equalisation busbar of the electrical installation . To use the Potential Equalisation Connector, connect the Potential Equalisation
Connector on the Workstation to the potential equalisation busbar of the electrical installation .

Patents

Covered under the following Patents: EP 1502612 (DE, CH, FR, GB, ES & IT), CA 2614658 . Design Patents: DE 49910883; FR 997137; and JP
1117998
1000DF00742 Issue 2
25/30