Alaris™ Gateway Workstation v1.2 Introduction Introduction The Alaris™ Gateway Workstation v1 .2 (hereinafter referred to as Workstation) has been designed as a modular system to provide a communications gateway between an Alaris Infusion Pump (hereinafter referred to as Pump) and any Patient Data Management System (PDMS), Patient Monitoring (PM) System, Hospital Information System (HIS) or Clinical Information System (CIS) that requires access to the infusion data retained within the Pump .
There are different software versions that can be ordered for each Primary Workstation . The software version of Workstation can be identified by the labelling on the front . Please contact local CareFusion representative for further information about Workstation versions, upgrade options and product availability .
Intended Use The Alaris Gateway Workstation v1 .2 is intended to be used to provide mounting, power and communications support to the Alaris Infusion Pumps range within the operating environment specified in this Directions For Use (DFU) . In such environments, the Workstation may be exposed to the following range of therapies: Fluid therapy, blood transfusions, parenteral feeding, drug therapy, chemotherapy, dialysis and anaesthesia .
Page 7
Workstation from a client PC . The Workstation Web Browser User Interface is validated on Windows Internet Explorer 10 . The software is provided under and is subject to a licence from CareFusion . 1000DF00742 Issue 2...
Alaris™ Gateway Workstation v1.2 Introduction Workstation Identification The Workstation can be confirmed by viewing the serial number label on the rear of the Workstation and checking the SKU is 80203UNSxx, where xx indicates the connectivity option . The two or three numbers at the end of the SKU identifies the Workstation configuration, see below for all available configurations .
It is important to ensure that you only refer to the most recent version of the Directions for Use and Technical Service Manual for your CareFusion products. These documents are referenced on www.carefusion.com. Copies can be obtained by contacting your local CareFusion representative.
Alaris™ Gateway Workstation v1.2 Controls and Indicators Controls and Indicators Controls Symbol Description ON/OFF Button - Press once to switch the Workstation on . Press and hold for two seconds to switch the Workstation off . In the event that the system needs to be reset, press and hold for at least four seconds, then press again to switch the Workstation on .
Alaris™ Gateway Workstation v1.2 Symbol Definitions Symbol Definitions Labelling Symbols Symbol Description Nurse call Connector RS232 Connector Auxiliary Connector Interface Device, General (Barcode Reader Connector) Ethernet Network Connector AC Inlet AC Outlet - Do Not Use This equipment contains an RF transmitter (where fitted) Fuse Rating Consult accompanying documents Potential Equalisation (PE) Connector...
Alaris™ Gateway Workstation v1.2 Features of the Workstation Features of the Workstation Modular Design The Workstation is a modular design . The Base Module comprises 3 MDI tiles with modules of 2 MDI tiles expanding the vertical configuration . Horizontal T-pieces of 2, 3 or 4 MDI tiles may be added to accommodate Pumps and fluid bag hangers, as required . The Workstation can only be modified and assembled by Qualified Service Personnel .
Page 13
Alaris™ Gateway Workstation v1.2 Features of the Workstation Configuration 80203UNSxx-235, shown mounted on an Alaris Trolley: Bag Supports Bag Support Rail Mounting Rail Infusion Line Tidies Mounting Rail Mounting Rail Trolley Note: Qualified Service Personnel are required to assemble the Trolley and to mount the Workstation on the Trolley . 1000DF00742 Issue 2 12/30...
Alaris™ Gateway Workstation v1.2 Features of the Workstation Infusion Line Tidies (where fitted) To assist in the routing of infusion sets and syringe extension sets from the Pumps to the patient, line tidies can be attached to the rear of the Workstation .
Alaris™ Gateway Workstation v1.2 Features of the Workstation Power Input The Workstation is powered from the AC supply through a standard IEC AC connector . When connected to the AC supply the AC Power indicator is illuminated . Both the Live and Neutral lines of the main supply are protected using fuses carried in a double fuse holder located on the AC inlet connector .
Alaris™ Gateway Workstation v1.2 Features of the Workstation System Fault Indication Continuous monitoring of the power distribution and communications system integrity is performed by the Workstation . In the event that a system fault occurs whilst in use, the System Fault Indicator will be illuminated accompanied by an audible tone . To avoid any possible interruption to the infusion, AC power to the Pumps will be maintained on the MDI tile should a system fault be detected .
Workstation; or MRI image distortion . This safe distance should be established in accordance with the manufacturers’ recommendations regarding electromagnetic interference (EMI) . For further information, please refer to the product technical service manual (TSM) . Alternatively, contact your local CareFusion representative for further guidance .
CareFusion does not provide any warranty for or endorsement on any Alaris Gateway Workstation that has been so modified or altered . CareFusion product warranty shall not apply in the event the Alaris Gateway Workstation has suffered damage or premature wear, or malfunctions or otherwise operates incorrectly, as a result of unauthorised modification or alteration of the Alaris Gateway Workstation .
Alaris™ Gateway Workstation v1.2 Operating Precautions Workstation Mounting Precautions • When mounting the Workstation to a mounting option other than pole or equipment rail, the following warnings must be observed: – Warning: Ensure the mounting option is capable of supporting a fully laden Workstation (see Product Specifications) prior to mounting .
Alaris™ Gateway Workstation v1.2 Operation of Workstation Operation of Workstation Switching On After initially switching on the Workstation, any services and applications running on the Workstation may take up to 90 seconds to become fully operational . 1 . Connect the AC power cord from the AC supply to the AC power inlet socket on the Workstation . 2 .
Alaris™ Gateway Workstation v1.2 Operation of Workstation Removing an Auxiliary Workstation If the primary Workstation cannot be switched off for whatever reason then the Pumps fitted to the auxiliary Workstation can be removed but the auxiliary Workstation must remain on. If the auxiliary Workstation is turned off or disconnected it may cause the primary Workstation to activate the System Fault Indicator.
Alaris™ Gateway Workstation v1.2 Data Communication Interfaces Data Communication Interfaces The user should be familiar with the data communications interfaces available on the Workstation before attempting to connect the Workstation to client / server systems . Erroneous connection of data communication cables will not damage the Workstation, but may cause the Workstation to operate incorrectly until the error is fixed .
Alaris™ Gateway Workstation v1.2 Barcode Reader (optional) Barcode Reader (optional) Overview The barcode reader (where fitted) allows the scanning and recognition of barcodes . Cable Connection Green LED Red LED Laser Aperture The Workstation must be powered down before attaching or disconnecting the barcode reader. Visual Indicators The barcode reader has two LED indicators (green and red) located on the top of the reader .
Alaris™ Gateway Workstation v1.2 Product Specifications Product Specifications Electrical Protection Against Electrical Shock: Class I Supply Voltage: 115-230V, ~50-60Hz Rating: 460VA (Maximum) Fuses: 2 x 4AH Time-Lag T Consult Qualified Service Personnel for replacement of fuses . AC Outlets: MDI Tile: 115-230V, ~50-60Hz, 60VA Protection against fluid ingress: IPX1 - Protected against vertically falling drops of water .
. Any use of battery packs that are not manufactured by CareFusion in the Alaris Gateway Workstation is at your sole risk, and CareFusion does not provide any warranty for or endorsement on any battery packs that are not manufactured by CareFusion .
. To discard electrical and electronic equipment, please contact local CareFusion affiliate office or distributor for further information . Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human health and the environment which could otherwise arise from inappropriate waste handling .
The Technical Service Manual (1000SM00015) is now available in electronic format on the World Wide Web at : www.carefusion.co.uk/alaris-technical/ A username and password are required to access our manuals . Please contact a local customer services representative to obtain login details .
Alaris™ Gateway Workstation v1.2 Service Contacts Service Contacts For service contact your local Affiliate Office or Distributor . CareFusion, CareFusion, CareFusion, CareFusion, PO Box 5527, Tullastr . 8-12 Döbrentei tér 1, Avda . São Miguel, 296 Atelier 14 Dubai, United Arab Emirates .