Datex-Ohmeda Aestiva 7900 SmartVent Technical Reference Manual page 131

Anesthesia ventilator
Hide thumbs Also See for Aestiva 7900 SmartVent:
Table of Contents

Advertisement

Message
Alarm type/
Priority
Inspiration
high
Stopped
(2), (3)
Internal
Minimum
Ventilator
Shutdown
Clock Too
(6)
Fast (7)
Internal
Minimum
Ventilator
Shutdown
Clock Too
(6)
Slow (7)
Invalid
User Alarm
Circuit
Module
Limit Task
User Alarm
Light Use
Loss of
User Alarm
Backup
Audio
Low Battery
User Alarm
Voltage
Low Drive
User Alarm
Gas Pres
Low O2
high
1006-0453-000
05/04
Enabling Criteria
Alarm Condition
Removal Criteria
Drive gas safety switch
engaged
Ventilator malfunction.
Ventilate manually.
Monitoring is not
reliable.
Ventilator malfunction.
Ventilate manually.
Monitoring is not
reliable.
The system does not
recognize the type of
circuit module installed.
Normally the system
used the ID tabs to
Identify circuits.
The system is running
on battery power.
No action is required. Turn off the light to save
power.
The audio alarm will not
Monitor system
sound for a CPU failure.
operation.
Voltage is <11.65V
Manually ventilate the
while using battery
patient to save power.
power.
The ventilator did not
Manually ventilate the
detect a rise in internal
patient.
pressure when the flow
valve opened.
In-range O
data
O
< low limit
2
2
available
Alarm condition becomes false or enabling criteria
becomes false.
5 Troubleshooting
Service Repair
Check the breathing circuit.
Check ventilator flow output.
Perform flow sensor calibration.
Perform the pressure limit switch test in the
service screen. If test fails, replace the SIB
board or SIB to CPU cabling
If problem persists, replace the CPU board.
Replace the CPU board.
Replace the CPU board.
Make sure the module is correctly installed.
Look for broken ID tabs or tape on the tabs.
If persists, check:
• circuit identification board
• harness to SIB
• SIB
• cable from SIB to CPU board
Replace the CPU board.
Make sure power is connected and circuit
breaker is closed. Check ventilator fuse.
Make sure that the appropriate gas supplies
(O
or air) are connected and pressurized.
2
Check O
limit. Is actual O
lower than limit
2
2
setting?
Verify O
calibration has been done correctly.
2
Does calibration pass?
If calibration does not pass, replace O
sensor.
If there is still a calibration problem, replace
the SIB board.
2
5-15

Advertisement

Table of Contents
loading

Table of Contents